Abstract:
If you are using ExamWRITER with the Zeiss i.Profiler equipment interface and it is not working, follow the steps below, provided by Carl Zeiss:
i.Scription Steps
- Open the i.Scription Software Settings window.
- Click Add New and record the i.Profiler Plus IP address.
- Select 1 from the Search for measurements from last drop-down menu.
- Select the disable authentication check box in the Authentication box.
- Do not modify the export or import directory paths. The export directory path should be C:\Documents and Settings\All Users\Application Data\Pontus\PMS_Export. The import directory path should be C:\Documents and Settings\All Users\Application Data\Pontus\PMS_Import.
ExamWRITER Steps
- Ensure that you are using ExamWRITER v9.5.12 or above.
- Open ExamWRITER.
- On the ExamWRITER main window, click Tools and select Equipment Integration Setup.
- Select the check box (Zeiss iScription Export, Zeiss iScription Import, Zeiss iProfilerPlus) next to the equipment that you are using.
- Select the location on the network where Zeiss saves the data file:
- Select a drive from the drop-down menu.
- Select the folder in which Zeiss saves the data file. The Zeiss iScription Export folder should be located at C:\Documents and Settings\All Users\Application Data\Pontus\PMS_Export. The Zeiss iScription Import folder should be located at C:\Documents and Settings\All Users\Application Data\Pontus\PMS_Import. The Zeiss iProfilerPlus folder should be located at C:\iProfiler (if this folder does not exist, create it now).
- Type the name of the i.Profiler Plus in the Computer Name